QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 Details

Hemodialysis is a medical procedure to remove fluid and waste products from the blood and to correct electrolyte imbalances. This is accomplished using a hemodialysis machine. In this process, body blood is filtered outside your body, cleaned, and then returned back. The dialysis machine itself can be thought of as an artificial kidney.Hemodialysis is used to treat both acute and chronic kidney failure.
QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is a Double Pump Blood Kidney Dialysis Machine. It is one of the the latest generation machines offered by QualiMed with advanced technology, therapy options, and monitoring features, while preserving unmatched user-friendliness.
So, What makes QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 distinguished?
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 offers easy installation
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 offers high precision
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is built with state of the art engineering
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is high quality and world class.
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is safe to use
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is user friendly
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 comes with easy upgradation and the best after sales support available 24/7
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 is applicable for chronic renal failure and other blood purification treatment.
- This device is specially designed, produced and sold for renal failure patients to receive haemodialysis, which is not allowed to be used for other purposes.
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 serves the patients of renal failure, kidney disease and uremia.
- It is also used for the patients with poisoning.
- This device should be used in medical units.
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 facilitates Pure ultrafiltration Therapy
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 facilitates Hemoperfusion Therapy
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 offers Plasma exchange Therapy
- Na+and UF profiling can be done by QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2.
- It is applicable for reducing the cases during dialysis treatment such as low/high pressure, muscle convulsion, headache, nausea, vomit etc
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 can compare dialysis parameters of the new single-pump SN HD system with a double-pump SN HD system available on the market.
- QualiMed Henle Series II of Hemodialysis Machine QMCHDHL2 consists of a large central tower with a metal stand with 4 hooks, positioned in an 'x', on the front right corner; outlet cord attached at the rear left of the machine and extends down to a three pronged plug; right rear has two hoses protruding; front of machine contains control panels, display panel and three hoses
- Touch screen was equipped and it can show Na and UF profiles; In addition, the functions ie. backup power supply, Kt/V,HCO3 adjustment, single needle, acetate treatment, pyrogen filter port and net connection etc are optional.
- Size & Weight Size: 380mm×400mm×1380mm(L*W*H)
- Area: 500 x 520 mm
- Weight: 88kg
- Power Supply AC220V, 50Hz / 60Hz,10A
- Input power: 1500W
- Back-up battery: 30 minutes (optional)
- Water input pressure: 0.15 MPa~0.6 MPa, 21.75 P.S.I.~ 87 P.S.I.
- Water input temperature: 10℃~30℃
- Working environment: temperature 10ºC ~ 30ºC at relative humidity of no more than 70%.
Dialysate
- Dialysate temperature: preset range 34.0℃~39.0℃
- Dialysate flux: 300~800 ml/min
- Dialysate concentration: 12.1 mS/cm ~ 16.0 mS/cm, ±0.1 mS/cm
- Dialysate mixing ratio: can set variety ratio.
- UF rate Flow range: 0 ml/h ~ 4000 ml/h
- Resolution ratio: 1ml
- Precision: ±30 ml/h
Extracorporeal Part
- Venous pressure: -180 mmHg ~ +600 mmHg, ±10 mmHg
- Arterial pressure: -380 mmHg ~ +400 mmHg, ±10 mmHg
- TMP pressure: -180 mmHg ~ +600 mmHg, ±20 mmHg
- Blood pump flow range: 20 ml/min ~ 400 ml/min(diameter:Ф6 mm)
- Spare pump flow range: 30 ml/min ~ 600 ml/min(diameter:Ф8 mm)
- Resolution ratio: 1 ml
- Precision: error range ±10ml or 10% of reading
Heparin Pump
- Syringe size: 20, 30, 50 ml
- Flow range: 0 ml/h ~ 10 ml/h
- Resolution ratio: 0.1ml
- Precision: ±5%
Sanitize
- Hot decalcification:
- Time: about 20 minutes
- Temperature: 30~60℃, 500ml/min.
- Chemical disinfection:
- Time: about 45 minutes
- Temperature: 30~40℃, 500ml/min.
- Heat disinfection:
- Time: about 60 minutes
- Temperature: >85℃, 300ml/min.
Storage Environment Storage temperature should be between5℃~40℃, at relative humidity of no more than 80%. Monitoring System
- Dialysate temperature: preset range34.0℃~39.0℃, ±0.5℃
- Blood leak detection: Photochromic
- Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate
- Bubble detection: ultrasonic
- Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow
- Conductivity: acoustic-optic, ±0.5%
Kt/V function
- The sufficiency of dialysis can be reflected by Kt/V.
- HCO3 adjustment function:
- Be applicable for sericous acid toxicosis patient.
- Be capable to calibrate the acid toxicosis status according to patient's state of illness.
- Backup power-supply:
- It can last for 15-30 minutes after power cut off.
- Pyrogen filter port
- Completely filtrating the remained bacteria and endotoxin.(optional)
- Net connection
- One computer can audit and control the treatment of maximum 20 units hemodialysis machine (optional)